On Flow Measurement F-0009

Required Straight Run to install flowmeters

To obtain specified accuracy, straight run both for upstream and downstream is needed. The required straight run is depeding on the type of flowmeter.
The recommended straight run by type is shown below.

Type of
Flowmeter
Tokyo Keiso
Model
Required straight run Remarks
Upstream Downstream
Variable Area
Flowmeter
R-100
AC
AM-1000
M
A
P etc.
Not
required
Not
required
Turbulence in large size flowmeter may result unstable float movementr.
Magnetic
Flowmeter
MAGMAX Series 5D* 2D Flow control valves are to be located downstream, if any.
Ultrasonic
Flowmeter
UCUF Not required Not
required
Flow control valves are to be located downstream, if any.
UL320
UL610P
UL3030K
10D 5D
DP type
Flowmeter
Orifice type O-010
O-100
O-1000
O-180(780)
20D 10D ---
Pitot Tube type HDT
CFW-1000, 2000
CDT
---
V Cone Flowmeter VC,VD,VM Not required 3D 2D upstream and 5D downstream is recommended in case that valve and other fittings than elbows are located.
Vortex
Flowmeter
VF-2000
VF-2200
VF-2300
VF-3000
VF-4000
VF-5000
KARUMAN ACE
10D 5D Flow control valves are to be located downstream, if any.
Wheel type
Flowmeter
W-100
W-200
W-300
W-400
W-800
W-2000
W-3000
MU-1000
TW-070
TW-080
TW-090
TW-100
Not required Not
required
The inner diameter of the process piping is to be larger than the flowmeter's diameter.
W-500
WX-2000
20D 5D Flow control valves are to be located downstream, if any.
Corioli Force
Mass flowmeter
MASSMAX7000
MASSMAX7100
Not required Not
required
---
Thermal Mass
flowmeter/
Conroller
TF
TC
HM
Not required Not
required
5D both upstream and
downstream for type
TF-1150,A1250,A1251,
A1551,A2251.
5D upstream 0D for
downstream for
TF-2421 to 2441.
Thermal
flowmeter
TH 20D* 5D 3D upstream 0D for downstream for TH-1600 and TH-1800 with flow strainer.

D: Inside diameter of piping. (5D: 5 times of Inside diameter of piping)

Recommended straight run in the table is subject to the condition that one 90 deg elbow is located upstream. Double elbows, pumps and valves upstream may require further straight run.