contact us

Technology

On Flow Measurement F-0002

Introduction of Capacitance detection type Compact
Electromagnetic Flowmeter [2]/[4]

Specification
Measurable Fluid Liquids
Meter Size 25, 40, 50, 80, 100 mm
Measuring Range Flow velocity
Min. 0 to 0.3 m/s
Max. 0 to 12 m/s
Liquid Conductivity Min. 0.05μS/cm (1μS/cm for Water)
Fluid Temperature - 25 to 100 °C
Fluid Pressure Meter size
25 to 80 mm : -0.1 to 4 MPa
100 mm : -0.1 to 1.6 MPa
Accuracy ± 0.5 % of reading (Flow velocity >= 1 m/s)
Wetted Part Material
Measuring tube : 25 mm : Zirconia Ceramic
40 to 100 mm : Alumina Ceramic
Earthing rings : Stainless steel(316SS), HastelloyB/C,
Titanium, etc.
Gasket for Earthing
rings :
Fluororesin
Housing Material Primary head : Stainless steel
Converter : Die-cast aluminum
Process Connection Wafer
Output / Input Current output : 4 - 20 mA DC or 0 - 20 mA DC
Pulse output : Open collector
Status output : Open collector
Control input : Voltage
Indicator 2-lines LCD
Power Supply 100, 120, 200, 240 V AC
24 V DC
Face to face Dimension Mass
Mass Meter Size (mm) 25 40 50 80 100
Face to face (mm) 69 94 114 164 214
Mass (kg) 3.9 4.7 5.2 7.7 11.1